MAINS OVERCURRENT PROTECTION:
Combined 2-pole circuit breaker/power switch.
PATIENT CONNECTIONS TO NEUTRAL: Less than10 µA typical.
POWER DISPLAY ACCURACY:
The greater of +/- 10% of display or +/- 3 watts rated load.
LINE REGULATION:
Power change < 1% /V for vline = min to max voltage
OPERATING CONDITIONS:
10°C TO 30°C, 95% RH non-condensing.
MOUNTING RESTRICTION:
2" (5CM ) clearance required on each side of unit for cooling.
COOLING:
Natural convection, conduction, and radiation. No fan.
PAD MONITOR:
SINGLE PAD MODE: Two wire continuity detector, typical threshold= 10 OHMS
DUAL PAD MODE: Two wire resistance monitor, typical acceptance range 10 to 150 OHMS, trip threshold typically 20% higher than the last monitor set point activation, visual indication of patient resistance.
ACTIVATION FREQUENCIES:
CUT = 520 HZ
COAG = 400 HZ
OPERATOR ERROR = 1710 HZ, pulsating
PLATE ALARM = 1000 HZ (65 DBA MINIMUM)
MACHINE FAULT = 1000 HZ
(45 DBA MIN. EXCEPT ALARMS ARE 65 DBA MIN)
HEIGHT: 6.25" (15.9) CM)
WIDTH: 15.9" (40.4 CM)
DEPTH: 22.75" (57.8 CM)
PARAMETERS OUTPUT NOMINAL AND MODES | ||||||
MODE | MAX POWER (WATTS) | RATED LOAD (OHMS) | TYPICAL CREST FACTOR | MAXIMUM OPEN CIRCUIT VOLTAGE (KHZ) | CARRIER REQ. (KHZ) | PULSE REPETITION FREQ (KHZ) |
MONOPOLAR | ||||||
300 | 300 | 1.7 - 1.9 | 2000 | 416.7 | CONTINUOUS | |
BLEND 1 CUT | 180 | 300 | 2.3 - 2.5 | 2000 | 416.7 | 20.0 |
BLEND 2 CUT | 120 | 300 | 2.9 - 3.2 | 2200 | 416.7 | 20.0 |
BLEND 3 CUT | 80 | 300 | 3.6 - 4.0 | 2400 | 416.7 | 20.0 |
STD COAG | 120 | 300 | 6.2 - 6.5 | 6500 | 540 | 20.0 |
SPRAY COAG | 80 | 500 | 8.0 - 9.0 | 1050 | 540 | 20.0 |
BIPOLAR | ||||||
CUT | 50 | 50 | 1.6 - 1.8 | 400 | 1000 | CONTINUOUS |
COAG | 50 | 50 | 1.5 - 12.0(BIPOLAR COAG CREST FACTOR DECREASES WITH POWER SETTING | 400 | 1050 | 20.0 |